User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: EDC HS-5-500:.1ê 5A Shunt DATE: 07-Jun-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 9 NUMBER OF LINES: 140 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A CONFIGURATION: Datron 1281 STANDARD: Leeds & Northrup 4223-B STANDARD: Thermo Electric MM II (1) STANDARD: Thermo Electric MM II (2) ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Leeds & Northrup 4223-B 1.004 OPBR Is the MM II's DTB number 25-30? 1.005 JMPF 1.008 1.006 STD Thermo Electric MM II (1) 1.007 JMP 1.009 1.008 STD Thermo Electric MM II (2) 1.009 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.010 DISP Connect the HI CURRENT OUTPUT from the Fluke 5725A 1.010 DISP to one of the large current terminals on the .001ê 1.010 DISP Leeds & Northrup Standard Resistor. 1.011 DISP Connect the other current terminal on the standard 1.011 DISP resistor to one of the current terminals on the UUT. 1.012 DISP Connect the other current terminal on the UUT 1.012 DISP to the LO CURRENT OUTPUT on the 5725A. 1.013 DISP Connect the potential terminals on the Leeds & Northrup 1.013 DISP resistor to the HI and LO front inputs on the 1281 DMM. 1.014 DISP Connect the potential terminals on the UUT to the 1.014 DISP rear CH B input on the 1281 DMM. 1.015 DISP Insert temperature probe into the center well 1.015 DISP of the standard resistor. 1.016 IEEE [@1281]*RST 1.017 IEEE [@1281]DCV AUTO;RESL7 1.018 5700 5A B1 S 2W 1.019 DISP Observe the Datron 1281, until the millivolt 1.019 DISP reading stabilizes. 1.020 CALL Standard Resistor Correction 1.021 MEME >3 1.022 HEAD {} 1.023 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.024 HEAD {º DC Calibration º} 1.025 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.026 HEAD ÍÍ DC Calibration ÍÍ 1.027 JMP 2.001 1.028 EVAL 2.001 IEEE [D3000] 2.002 IEEE [@1281]RDG?[I] 2.003 MEME >1 2.004 IEEE [@1281]INPUT CH_B 2.005 IEEE [@1281][D1000]RDG?[I] 2.006 MEME >2 2.007 5700 * S 2.008 ACC .100000Z 0.0100% 2.009 MATH m[1]=m[1]/m[3] 2.010 MATH m[2]=m[2]/m[1] 2.011 MATH mem=.1 2.012 MATH mem1= m[2] # 2.013 MATH mem1=mem1*1000 2.013 MEMC Z 0.100000% ÿ5A 3.001 IEEE [@1281]*RST 3.002 5700 3A B1 S 2W 3.003 IEEE [@1281]DCV AUTO;RESL7 3.004 IEEE [D3000] 3.005 IEEE [@1281]RDG?[I] 3.006 MEME >1 3.007 IEEE [@1281]INPUT CH_B 3.008 IEEE [@1281][D1000]RDG?[I] 3.009 MEME >2 3.010 5700 * S 3.011 ACC .100000Z 0.0100% 3.012 MATH m[1]=m[1]/m[3] 3.013 MATH m[2]=m[2]/m[1] 3.014 MATH mem=.1 3.015 MATH mem1= m[2] # 3.016 MATH mem1=mem1*1000 3.016 MEMC Z 0.100000% ÿ3A 4.001 IEEE [@1281]*RST 4.002 5700 1A B1 S 2W 4.003 IEEE [@1281]DCV AUTO;RESL7 4.004 IEEE [D3000] 4.005 IEEE [@1281]RDG?[I] 4.006 MEME >1 4.007 IEEE [@1281]INPUT CH_B 4.008 IEEE [@1281][D1000]RDG?[I] 4.009 MEME >2 4.010 5700 * S 4.011 ACC .100000Z 0.0100% 4.012 MATH m[1]=m[1]/m[3] 4.013 MATH m[2]=m[2]/m[1] 4.014 MATH mem=.1 4.015 MATH mem1= m[2] # 4.016 MATH mem1=mem1*1000 4.016 MEMC Z 0.100000% ÿ1A 5.001 IEEE [@1281]*RST 5.002 HEAD {} 5.003 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 5.004 HEAD {º AC Calibration º} 5.005 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 5.006 HEAD ÍÍ AC Calibration ÍÍ 5.007 JMP 6.001 5.008 EVAL 6.001 DISP Disconnect the test setup. 6.002 DISP Connect the UUT current terminals directly to the 6.002 DISP Fluke 5725A CURRENT OUTPUTS. 6.003 DISP Connect the UUT potential terminals to the Datron 1281 6.003 DISP HI/LO INPUTS. 6.004 IEEE [@1281]*RST 6.005 IEEE [@1281]ACV AUTO, FILT40HZ 6.006 5700 5A 40H B1 S 2W 6.007 IEEE [@1281][T0]RDG?[I] 6.008 ACC .10000Z 0.06% 6.009 MATH MEM=MEM/5 6.010 MEME 6.011 MEMC Z 0.5% 40H 7.001 IEEE [@1281]ACV AUTO, FILT100HZ 7.002 5700 5A 100H B1 S 2W 7.003 IEEE [@1281][T0]RDG?[I] 7.004 ACC .10000Z 0.06% 7.005 MATH MEM=MEM/5 7.006 MEME 7.007 MEMC Z 0.5% 100H 8.001 5700 5A 400H B1 S 2W 8.002 IEEE [@1281][T0]RDG?[I] 8.003 ACC .10000Z 0.06% 8.004 MATH MEM=MEM/5 8.005 MEME 8.006 MEMC Z 0.5% 400H 9.001 HEAD {} 9.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 9.002 DISP {º THIS COMPLETES THE VERIFICATION º} 9.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 9.003 END